Product Information

Picarro G5131-i Analyzer

For the precise measurement of N2O Concentration and Isotopes analysis for δ15N, δ15Nα, δ15Nβ, δ18O

With the G5131-i next to the N2O concentration also the isotopic composition of N2O (δ15N including site-specific isotopomers and δ18O) can be measured continuously and precisely. The device is so designed that the best results are obtained at atmospheric concentrations. As almost all of the Picarro devices it is suitable for field sites as well as for the laboratory use as it inter alia allows for a cryogen-free continuous operation.

 

 

 

 

 

Technical Specifications

Technical Specifications

Precision
1-σ
10 min avg
Precision
1-σ
100 sec avg
Concentration
Range 


(ppb N2O in Air)
Max Drift
(over 24hrs, peak to peak, 1hr average)
N2O (Concentration) Typical: < 0.04 ppb
Guaranteed:            < 0.05 ppb
Typical: < 0.07 ppb
Guaranteed:             < 0.1 ppb
300 - 1500< 0.1 ppb Typical
δ15N, δ15Nα, δ15 Typical: 0.3 ‰
Guaranteed:             < 0.5 ‰
Typical: 0.7 ‰
Guaranteed:               < 1‰
300 - 1500< 2 ‰ Typical
δ18OGuaranteed:             < 0.7 ‰Guaranteed:             < 1‰
(300 sec  avg)
300 - 1500< 2 ‰ Typical

 

 

Order Numbers + Accessories

 

Ordering Details:

Art.-Nr.:  196382 For the measurement of N2O Concentration and Isotopes analysis for δ15N, δ15Nα, δ15Nβ, δ18O

 

 

Downloads

Hauptsitz
Müncheberg
Tel.: +49 (0) 33432 - 89 575
Fax: +49 (0) 33432 - 89 573

Niederlassung Süd
Tel.: +49 (0) 8161 - 23 46 441